胡其会, 张鸣远, 李景银, 等. 轴流式血泵水动力特性和生物相容性的数值模拟[J]. 西安交通大学学报, 2011,45(8):118-122.
Numerical Simulation for Hydrodynamics and Biocompatibility in an Axial Blood Pump[J]. 2011, 45(8): 118-122.
为了准确评价自行设计的轴流式血泵的水力性能和生物相容性
将一种黏弹性流体k-ε模型应用于该血泵内紊流流动特性的数值模拟
分析了轴流泵的泵壳与转子间狭缝中以及叶轮流道内的静压和切应力的分布规律
并将溶血指标计算方法与此黏弹性紊流模型结合
计算了黏弹性流体为工质时血泵的溶血指标
并与水为工质时的计算结果进行了对比.结果表明:在相同转速和流量条件下
质量分数为0.06%的黄原胶溶液为工质时的静压压升高于以水为工质时的相应值
质量分数为39%的甘油溶液为工质时静压压升最不明显; 甘油溶液的雷诺应力和黏性切应力均高于水和黄原胶溶液的相应值; 黏弹性流体为工质时泵的溶血指标比水和甘油溶液的低
且数值处于同一数量级上.该模型能较为准确地预测血泵内黏弹性流体的紊流流动特性
但血泵的结构需进一步优化和改善.
To accurately evaluate the hydraulic performance and biocompatibility of an axial blood pump designed and tested
a k-ε model for viscoelastic fluids is employed to predict the turbulent flow characteristics in the pump. The distributions of static pressure and wall shear stress in the clearance gap between the rotor and the pump shell and in the main flow passage are analyzed. A numerical method for estimating hemolysis level in blood pump is used to calculate the hemolysis indexes of viscoelastic fluids and water. The simulation shows that the static pressure for 0.06%(mass fraction)xanthan gum solution is higher than that for water at same flow rate and same rotating speed
and the static pressure of 39%(mass fraction)glycerin solution is the lowest among them all. The Reynolds shear stress and the viscous shear stress for glycerin solution are higher than those for water and xanthan gum solution. The hemolysis index of viscoelastic fluids is lower than those of water and glycerin solution
and the hemolysis indexes of them all are in the same order of magnitude. This turbulence model can predict the turbulent flow characteristics of viscoelastic fluids in the pump accurately
and the structure of the axial blood pump is expected to be further optimized and improved.
朱宪然. 一种轴流式血泵的设计及其非牛顿流体流动特性研究[D]. 西安: 西安交通大学, 2007.
FRAZIER O H, TIMOTHY J M, JARVIK R K, et al. Research and development of an implantable, axial-flow left ventricular assist device: the jarvik 2000 heart [J]. Ann Thorac Surg, 2001,71:S125-S132.
GEORGE P N, DEBORAH M, SUELLEN L, et al. Development and clinical application of the MicroMed DeBakey VAD [J]. Current Opinion in Cardiology,2000,15(3):166-171.
PINHO F T. A GNF framework for turbulent flow models of drag reducing fluids and proposal for a k-ε type closure [J]. Journal of Non-Newtonian Fluid Mechanics, 2003,114(2/3): 149-184.
CRUZ D O A, PINHO F T. Turbulent pipe flow predictions with a low Reynolds number k-ε model for drag reducing fluids [J]. Journal of Non-Newtonian Fluid Mechanics, 2003,114(2/3): 109-148.
胡其会,张鸣远,杨万英. 轴流式血泵黏弹性流体水动力特性数值模拟[J]. 西安交通大学学报,2010,44(11):28-32.
HU Qihui, ZHANG Mingyuan, YANG Wanying. Numerical simulation on viscoelastic fluids hydrodynamics in an axial blood pump[J]. Journal of Xi'an Jiaotong University, 2010, 44(11): 28-32.
OLSEN D B. The history of continuous-flow blood pumps[J]. Artificial Organs, 2000, 24(6): 401-404.
WURZINGER L J, OPITZ R. Mechanical blood trauma: an overview [J]. Angeiologie,1986, 38:81-97.
BLUDSZUWEIT C. Model for a general mechanical blood damage prediction [J]. Artificial Organs,1995,19(7):583-589.
BLUDSZUWEIT C. Three-dimensional numerical prediction of stress loading of blood particles in a centrifugal pump [J]. Artificial Organs, 1995,19(7):590-596.
GOUBERGRITS L, AFFELD K. Numerical estimation of blood damage in artificial organs [J]. Artificial Organs, 2004,28(5):499-507.
朱宪然,张鸣远,刘昊南,等.轴流式血泵内部流场和生物相容性的数值分析[J].西安交通大学学报,2006,40(11):1290-1294.
ZHU Xianran, ZHANG Mingyuan, LIU Haonan, et al. Numerical investigation on flow field and biocompatibility of an axial blood pump[J]. Journal of Xi'an Jiaotong University, 2006, 40(11): 1290-1294.
YANO T, SEKINE K, MITOH S, et al. An estimation method of hemolysis within an axial flow blood pump by computational fluid dynamics analysis [J]. Artificial Organs, 2003, 27(10): 920-925.